Safety Practices . 23
5. Replace the fuse holder by aligning the lug at the bottom of the fuse holder with the
slot in the socket and then pushing the holder into place.
You can hear a click as each lug snaps into place.
NOTE: If you use the correct fuses but the instrument still does not work correctly, or the
fuses blow repeatedly, contact your PerkinElmer Service Engineer.