The petMAP+II monitor is designed to conform to Electromagnetic Compatibility (EMC)
standard EN 60601-l -2, 2007 and will operate accurately in conjunction with other medical equipment
which also meets this requirement. To avoid interference problems affecting the Monitor, do not use
the Monitor in the presence of equipment which does not conform to these specifications.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE
This product has been tested and found to comply with the limits for a Class B digital device, pursuant to
part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation. The product generates, uses and can radiate radiofrequency
energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur
in a particular installation. If the product does cause harmful interference to radio or television
reception, which can be determined by turning the product on and off, the user is encouraged to try to
correct the interference by one or more of the following measures:
• Re-orient or relocate the receiving antenna.
• Increase the separation between the product and receiver.
• Consult the dealer or an experienced radio/TV technician for help.
The petMAP+II is not intended for use on patients being imaged with an MRI device since the
petMAP device contains magnetically active materials and could result in injury if used too close to an
operating MRI.
Inaccurate readings may result when an electrosurgical unit (ESU) is used while monitoring with
the petMAP+II. If this is suspected, discontinue use of the petMAP+II while the (ESU) is in use.
Do not gas sterilize or autoclave the device.