SPECIFICATIONS
For Veterinary Use Only
Method of BP Measurement: Oscillometric. Accuracy optimized for species (dog/cat) and blood
pressure cuff site (forearm, tail, hind foot).
Parameters Measured:
BP: Systolic, Diastolic, Mean Arterial Pressure (MAP)
Heart Rate, both by BP cuff and SpO2 (if configured)
Sp02
Temperature. Esophageal or rectal probes are available
Blood Pressure Ranges:
Systolic: 40 265 mmHg
Diastolic: 25 195 mmHg
MAP: 30 200 mmHg
Heart Rate Range from BP: 40 240 BPM, +/- 5% or 3 BPM, whichever is greater.
Heart Rate Range from SpO2: 40 240 BPM, +/- 5%
SpO2 (if equipped):
SpO2 displays a value within the range 40 100%, display resolution: increments of 1%
SpO2 accuracy: +/- 3% at 80-100%; <80%, undefined.
Temperature
Range: 2C 46°C, display resolution of .1°C (84°F - 115°F)
Accuracy: +/- .2°C when connected to a YSI compatible thermistor probe.
Power: 4 AA batteries. Supplied with Rechargeable NiMH. Capable of using alkaline or lithium (these
batteries are not rechargeable and charging should not be attempted as a dangerous condition may
result. See Warnings).
Battery Life: NiMH batteries typically can be full-discharge/full-charge cycled several hundred times.
Operating Environment:
Temperature: 10°C 40°C (50°F 104°F)
Humidity: 15% - 85%, non-condensing
Altitude: -500 feet (152 meters) below sea level to +8000 feet (2438 meters) above sea level
Storage Temperature: -20°C to 55°C (-4°F to 131°F)
Dimensions: 5.97”H x 3.15”W x 1.26”D
Weight : 1 pound (0.46 Kg, approx, with NiMH batteries)