16
Electromagnetic Immunity
The petMAP+II is intended for use in the electromagnetic environment specified below. The customer or user of the
petMAP+II should assure that it is used in such an environment.
Immunity test
IEC60601
test level
Compliance Level Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part of
the device, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150kHz to 80 MHz
3 Vrms
d =(1.17)
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d =(1.17) 80 - 800MHz
d =(2.33) 800Mhz - 2.5GHz
where P is the maximum output power rating of the
transmitter in watts (W) and is the recommended
separation distance in meters (m). Field strengths
from fixed RF transmitters, as determined by an
electromagnetic site survey a , should be less than
the compliance level in each frequency range b.
Interference may occur in the vicinity of equipment
marked with the
following symbol: