Operating Instructions
Pharmaceutical Refrigerator
MPR-514

Questions and Answers

  • A
    Anthony WhiteAug 8, 2025
    What to do if Phcbi MPR-514R is not cold enough?
    • A
      Ashley PorterAug 8, 2025
      If your Phcbi Refrigerator's chamber isn't getting cold enough, several factors could be the cause. Start by reducing the amount of items inside, especially if you've recently added a large amount or warm products. Ensure you minimize how often the door is opened and that the chamber temperature isn't set too high. Check that the unit isn't in direct sunlight and is installed according to the operating instructions, with proper ventilation and no nearby heat sources. Also, ensure the ambient temperature is between -5 oC and +35 oC. Make sure the air exhaust vent is clear, the access port is covered with insulation and rubber caps, and there are no foreign substances between the door gaskets. If the door seal is damaged, contact your sales representative or agent for replacement.
  • T
    Tammy EvansAug 12, 2025
    What to do if my Phcbi Refrigerator is noisy?
    • S
      stonearthurAug 13, 2025
      If your Phcbi Refrigerator is making excessive noise, several factors could be the cause. Ensure the floor is sturdy and the installation site is level. Adjust the refrigerator to be level and make sure the cabinet isn't touching the surrounding wall. Additionally, check that the mounting plate or unit cover is correctly reinstalled after cleaning the evaporating tray.
  • M
    Mark GarciaAug 16, 2025
    What to do if nothing operates even when plugged in Phcbi MPR-514R Refrigerator?
    • C
      Christina NicholsAug 17, 2025
      If your Phcbi Refrigerator isn't operating even when plugged in, start by checking the power supply connection to ensure it's properly connected. Verify that the power supply has sufficient capacity and voltage. Check for a power failure and reset the circuit breaker if it has been activated. Finally, replace the fuse on the supply circuit if it's blown.
  • L
    luisdeanAug 21, 2025
    What to do if alarm device is activated during operation of Phcbi MPR-514R?
    • M
      Marvin DiazAug 21, 2025
      If the alarm is activated during operation of your Phcbi Refrigerator, check the power supply connection and ensure it has sufficient capacity and voltage. Also, check for a power failure and reset the circuit breaker if necessary. Replace the fuse if it's blown. Make sure the chamber temperature setting is correct, the door is closed, and no high-temperature items are inside. If the door was kept open, close it.
  • Y
    yweberAug 26, 2025
    Why alarm device is activated at start-up Phcbi Refrigerator?
    • M
      Meredith Powell MDAug 26, 2025
      The alarm status is kept until the chamber temperature reaches the set temperature.
  • C
    clinton16Aug 29, 2025
    How to unlock Phcbi MPR-514R key operation?
    • M
      Maria HessAug 29, 2025
      If no key operation is available on your Phcbi Refrigerator, the key lock may be set to ON (L 1). To resolve this, set the key lock to OFF (L 0).
  • K
    Krista HorneSep 3, 2025
    What to do if Phcbi MPR-514R Refrigerator returns to temp display mode?
    • T
      tyatesSep 3, 2025
      The setting mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation (auto-return function).

Phcbi MPR-514R Series Specifications

General IconGeneral
Voltage220-240V
Frequency50/60 Hz
Digital DisplayYes
ControllerMicroprocessor
Defrost SystemAutomatic
CategoryRefrigerator
CertificationCE