SECTION 1: TECHNICAL SPECIFICATIONS
Input
110-120 volts, 50 / 60Hz, single phase, 1.8 kVA, 16 amperes
OR
220-240 volts, 50 / 60Hz, single phase, 1.8 kVA, 8 amperes
Output Rating
0 -10 kilovolts, 300 mA
0 -50 kilovolts, 60 mA
Duty Cycle
5 MINUTES ON / 15 MINUTES OFF at 3 kVA
15 MINUTES ON / 1 HOUR OFF at 2 kVA
1 hour ON / 1 HOUR OFF at 1.5 kVA
Type of Cooling
ONAN (Oil-Natural, Air-Natural)
Distortion
Less than 5 percent at full load
Impedance
Less than 10 percent at rated current
Operating Ambient Temperature
0-40 degrees Celsius
Output Termination
Toroid
Metering
Output Voltmeter: Display: 3 ½ digit LCD
Accuracy: ± (0.8% of Reading + 0.2% of Range)
Ranges: 0 to 10kV / 50 kV
Output Currentmeter: Display: 3 ½ digit LCD
Accuracy: ± (0.8% of Reading + 0.2% of Range)
Ranges: 0 to 200µA / 2mA / 20mA / 300mA
Sizes and Weights
Control Section: 16.5" (419 mm) W x 15.25" (388 mm) D x 11.25" (286 mm) H
Weight: 29 pounds (13 kg) Add 3 pounds (1.4kg) for 220V
HV Transformer: 14” (356 mm) W x 14” (356 mm) D x 16.88” (429 mm) H
Weight: 80 pounds (36 kg)