EasyManua.ls Logo

Phenix Technologies 6CP50/10-3 - Page 7

Default Icon
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6CP50/10-3
3-3
CONTROLS AND INDICATORS
Control Panel (cont’d)
14. Measurement Divider. Setting of this switch determines the voltage measurement divider ratio. This
switch must be set to match the actual high voltage output connection.
WARNING!
If a test specimen intended to be tested at 25kV is connected to the 50kV output and the Measurement
Divider switch is set to 10kV, the actual voltage applied to the specimen will be approximately 5 times the
indicated voltage on the panel meter and an over-voltage may be applied to the test specimen.
15. CM Range. Turn to select the desired current measurement range.
16. Start/Reset. Press to start the timer after test voltage has been reached. Press again to reset the timer to
the original setting.
17. HV Off Bypass. Press to allow high voltage to remain on after timer has expired. Indicator lights when
depressed.
18. Timer. Select desired test time using buttons on face of timer. See Attachment