13 Fetal Heart Rate Alarms
128
Changing Alarm Delays
You can change the alarm delays in Configuration Mode.
1 Connect either an ultrasound or a DECG transducer to a free socket on the monitor.
2 Enter the Setup Menu for a connected FHR measurement.
3 To change the high alarm limit delay time, select High Delay and select the delay time (in seconds)
from the pop-up list.
4 To change the low alarm limit delay time in seconds, select Low Delay and select the delay time (in
seconds) from the pop-up list.
Changing Signal Loss Delay
The signal loss delay is the configurable delay before an INOP. You can change the delay in
Configuration mode:
1 Connect either an ultrasound or a DECG transducer to a free socket on the monitor.
2 Enter the Setup Menu for a connected FHR measurement.
3 Select SignalLoss Delay and select the signal loss INOP delay time (in seconds) from the pop-up
list.