6 Patient Alarms and INOPs
76
Technical Alarm Messages (INOPs)
INOP Message, Indication Source What to do
Batt EMPTY
INOP tone, battery LED flashes
during this INOP, alarms cannot
be paused or switched off.
Battery The estimated remaining battery-powered operating time is less
than 10 minutes. Connect the monitor to AC immediately.
If the condition persists and the monitor is not connected to
mains power, this INOP is re-issued two minutes after you
acknowledge it.
Batt INCOMPAT.
INOP tone
Battery The battery cannot be used with this monitor. Replace with the
correct battery.
Batt LOW
INOP tone
Battery The estimated battery-powered operating time remaining is less
than 20 minutes.
Batt MALFUNCTION
INOP tone, battery LED flashes
during this INOP, alarms cannot
be paused or switched off unless
the monitor is connected to
mains power.
Battery The monitor cannot determine the battery status. If this INOP
persists, replace the faulty battery. If the condition persists and
the monitor is not connected to mains power, this INOP is
reissued two minutes after you acknowledge it.
Batt MISSING Battery No battery found in FM20/30. Contact your service personnel.
BUS MASTER MALFUNC Monitor There is a problem with the monitor's hardware. Contact your
service personnel.
CHARGER MALFUNC.
INOP tone, battery LED may
flash
Battery There is a problem with the battery charger in the monitor.
Connect the monitor to mains power and contact your service
personnel.
CHECK BATT TEMP Battery The temperature of the battery is too high. Check that the
monitor is not exposed to heat.
Check Flex Texts
INOP tone
Monitor If this INOP appears, check the monitor and patient settings
before you resume monitoring. If the settings are unexpected,
there may be a problem with the monitor software. Contact your
service personnel.
Check Monitor Func
INOP tone
Monitor A potential internal problem with the monitor has been
detected. Contact your service personnel.
Check Keyboard
INOP tone
Monitor Perform a visual and functional check of the keyboard. Contact
your service personnel.
Check Mouse Device
INOP tone
Monitor Perform a visual and functional check of the mouse input
device. Contact your service personnel.
CHECK PAPER
INOP tone
Recorder Check that there is no paper jam, that the print drawer is
properly shut, that the paper is loaded with the grid facing
upwards, and that the correct Philips paper is being used.
Check Settings
INOP tone
Monitor If this INOP appears, check the monitor and patient settings
before you resume monitoring. If the settings are unexpected,
there may be a problem with the monitor software. Contact your
service personnel.