3 Basic Operation
54
Switching the Monitor to Standby
To switch the monitor to Standby:
Either
Select the
Monitor Standby SmartKey .
Or
1 Enter the Main Setup menu using the SmartKey .
2 Select Monitor Standby.
Pressing any key or selecting any field on the screen will resume monitoring.
After Monitoring
1 Discharge the patient.
2 Remove the transducer from the patient and, using a soft tissue, remove any gel from it. Then
clean the transducer.
3 Tear off the paper at the fold. To avoid misalignment of the recorder mechanism, NEVER pull on
the paper to advance it, or try to tear it other than at a fold (unless using the paper guide with the
FM40/FM50).
4 Switch off the monitor.
Disconnecting from Power
FM20/30 To disconnect the monitor from AC power, switch the monitor off using the On/Off switch located
on the right side of the device, or unplug the power cord from the AC mains socket.
FM20/30
(with
Battery
Option)
To disconnect from Battery power, switch the monitor off using the On/Standby switch located on
the right side of the device.
FM40/50
and FM20/
30 with
Battery
Option
The On/Standby switch does not disconnect the monitor from the AC power source. To disconnect,
unplug the power cord from the AC mains socket. Note that if the power cord is unplugged from the
AC mains socket before the monitor is put into Standby, a beeper is activated. The beeper warns you if
the monitor is accidentally disconnected from AC mains.