26 Specifications and Standards Compliance
200
SpO
2
Alarm Specifications Range Adjustment Delay
SpO
2
50 to 100% 1% steps (0, 1, 2, 3,... 30) +
4 seconds
Desat 50 to Low alarm limit 1% steps
Pulse 30 to 300 bpm 1 bpm steps (30 to 40 bpm)
5 bpm steps (40 to 300 bpm)
max. 14 seconds
Tachycardia Difference to high limit 0 to 50
bpm
5 bpm steps max. 14 seconds
Clamping at 150 to 300 bpm 5 bpm steps
Bradycardia Difference to low limit 0 to 50 bpm 5 bpm steps max. 14 seconds
Clamping at 30 to 100 bpm 5 bpm steps
SpO
2
Factory Default Settings
Desat Alarm Limit 80
Low Alarm Limit 90
High Alarm Limit 100
Desat Alarm Limit Delay 20 seconds
Low Alarm Limit Delay 10 seconds
High Alarm Limit Delay 10 seconds
Averaging Time 10 seconds
NBP Alarm Suppression On
Parameter Alarms On/Off On
Color Cyan
Pulse Settings
Pulse Alarms On/Off On
Pulse High Limit 120 bpm
Pulse Low Limit 50 bpm
Bradycardia: Difference to Low Limit 20 bpm
Bradycardia: Clamp 40 bpm
Tachycardia: Difference to High Limit 20 bpm
Tachycardia: Clamp 200 bpm