When power is lost - no power is available from the AC power source or from the battery - before the
monitor is put into Standby, a beeper will sound. The tone can be silenced by pressing the On/Standby
switch.
ESU, MRI and Defibrillation
WARNING
The fetal/maternal monitors are NOT intended for use during defibrillation, electro-surgery, or MRI.
Remove all transducers, sensors, and accessories before performing electro-surgery, defibrillation, or
MRI, otherwise harm to the patient or the user can result.
Cardiac Pacemakers and Electrical Stimulators
WARNING
The fetal/maternal monitors are NOT intended for use for ECG measurements on patients connected
to electrical stimulators or with cardiac pacemakers.
Fast Transients/Bursts
The equipment will return to the previous operating mode within 10 seconds without loss of any
stored data.