5 Alarms
70
then audible latching can only be set to Red or Off. The following table shows the possible
combinations for latching settings:
Alarm Latching Behavior
All INOPs except the "unplugged" INOPs are non-latching.
Possible Combinations for Alarm Latching Settings
Visual Latching Setting Audible Latching Setting
Red and Yellow Red and Yellow
Red and Yellow Red
Red and Yellow Off
Red Red
Red Off
Off Off
Alarm Condition Red and Yellow Measurement Alarms
Acknowledgment Presence Non-latching alarms Visual and audible
latching
Visual latching,
audible non-latching
Alarm has not been
acknowledged.
Alarm condition still
present.
Alarm tone on. Alarm
message.
Alarm tone on. Alarm
message. Flashing numerics.
Alarm tone on. Alarm
message. Flashing numerics.
Alarm condition no
longer present.
All audible and visual
alarm indicators
automatically stop.
Alarm tone on.
Alarm message.
Flashing numerics.
Alarm message. Flashing
numerics.
Audible alarm indicators
automatically stop.
Alarm has been
acknowledged.
Alarm condition still
present.
Alarm tone off. Alarm
message.
Alarm tone off. Alarm
message. Flashing numerics.
Alarm tone off. Alarm
message. Flashing numerics.
Alarm condition no
longer present.
Audible and visual alarm
indicators automatically
stop.
Audible and visual alarm
indicators automatically
stop.
Audible and visual alarm
indicators automatically
stop.