EasyManua.ls Logo

Philips DreamStation - Disconnecting the Tubing

Philips DreamStation
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
4. Carefully replace the water tank (1) into the humidier and close the humidier lid (2) until it snaps shut.
Warning: When installing the tank, do not allow any water to spill into the humidier or therapy device.
Caution: Do not move the humidier while the water tank has water in it.
5. Supply power to your therapy device. Refer to the manual included with your therapy device.
6. Put on your mask assembly. Refer to the instructions supplied with the mask.
7. Turn on the airow on your device and begin therapy. Refer to the manual included with your therapy device.
Note: If you are having trouble with your mask, refer to the instructions included with the mask.
8. Refer to your therapy device manual for complete instructions on how to adjust both the Heated Humidier and
the optional Heated Tubing settings to achieve the desired humidity.
Caution: Do not turn the humidier on without the water tank installed. The humidier setting must remain “off”
if there is no water in the water tank.
9. Refer to the manual included with your therapy device for instructions to turn off therapy.
DisconnectingtheTubing
1. To remove the heated tubing, press in the tabs (1) on the side of the tubing connector and pull the tubing away
from the outlet port.
2. If you are using a standard tube (not shown) instead of a heated tube, simply pull the tubing away from the outlet
port.

Other manuals for Philips DreamStation

Related product manuals