DisconnectingtheTherapyDevice
Caution: To avoid spilling, do not disconnect the humidier from the therapy device with water in the tank.
Remove the water tank from the humidier before disconnecting the therapy device.
1. Disconnect power to the therapy device.
2. Pick up the system.
3. Place one hand on the therapy device and the other on the humidier.
4. Press the humidier release button (1) and pull apart to separate.
CheckingtheHumidierLidSeal
Under normal use, the humidier lid seal should not require any maintenance or replacement. The seal may be
cleaned as needed by wiping it with a damp cloth. If necessary, the humidier lid seal may be removed for further
cleaning. Gently peel the seal from the humidier lid and clean it in a solution of warm water and a mild liquid
dishwashing detergent. Rinse with clean water. Wipe completely on both sides. Allow the seal to air dry. Inspect
the seal for damage. If the humidier lid seal show signs of wear or damage, contact your home care provider for a
replacement.
To install or reseat your humidier lid seal, fully open the humidier lid. Position the seal (1) against the inside of the
lid so the seal’s center hole aligns with the humidier outlet port. Conrm that the seal is positioned so the wire
channel (2) in the seal is below the humidier outlet port (3).
Note: The seal only ts properly in one orientation.
With the seal loosely in place, start at the bottom (1) and gently press the edges of the seal into the channel in the lid
of the humidier. Continue sliding your ngers all around the rectangular perimeter of the seal until the outer edge
is completely seated. Next, press the seal around the humidier outlet port (2) until the center of the seal is fully
seated. Finally, go back and run your ngers around the rectangular perimeter of the humidier lid seal once more to
conrm it has not become dislodged.