5
PERFORMANCE
CHECK
AND
ADJUSTMENT
PROCEDURE
REQUIREMENTS
Check
all
positions
of
the
MTB
TIME/DIV
switch.
Marker
input
and
MTB
TIME/DIV
0,2
us
0,5
us
1yus
2us
5
us
10
us
20
us
50
us
0,1
ms
0,2
ms
0,5
ms
ims
2ms
5
ms
10
ms
20
ms
50
ms
0,15
0,2s
0,5s
5.3.31
Sweep
length
adjustment
TEST
EQUIPMENT
_
Time-marker
generator
SETTING
Vertical
mode
-
A
Horizontal
mode
-
MTB
TIME/DIV
-
0,1
ms
maximum
coefficient
error
X
MAGN
x10
measured
over
central
8
div
X
MAGN
x1
+3%
+3%
+3%
+3%
+3%
+3%
+3
%
+3
%
+3%
+3
%
+3%
+3
%
+3%
+3%
+3%
+3%
+3%
+3%
+3%
+3%
+5
%
+5%
+5%
+5
%
+5
%-
+5%
+5%
+5
%
+5%
+5
%
+5
%
+5%
+5
%
+5%
+5%
+5
%
+5%
+5%
+5%
+5%
Apply
a
0,1
ms
marker
pulse
to
channel
A.
ADJUSTMENT
Adjust
VR614
(HOR
PCB)
for
a
sweep
length
of
11,5
divisions.
§.3.32
x10
MAG
time
adjustment
TEST
EQUIPMENT
_
Time-marker
generator
SETTING
Vertical
input
-
A
Horizontal
input
-
MTB
TIME/DIV
-
1
ms
X
position
switch
at
x10
Apply
a
0,1
ms
marker
pulse
to
channel
A.
ADJUSTMENT
Adjust
VR807
(HOR
PCB)
for
coincidence
of
the
marker
pulses
with
the
graticule
lines.
5.3
PERFORMANCE
CHECK
AND
ADJUSTMENT
PROCEDURE
5-13