Instructions for Use | Alarms and System Messages
Alarms and System Messages 45
High-priority System Alarms
VENTILATOR INOPERATIVE
The system self-test indicates a failure or malfunction of a component that causes therapy to stop or
not meet essential performance criteria.
Contact Philips Respironics Customer Service
performance
during active
Therapy is stopped and both audible and visual alarms are continuous. Depending on the systems
impacted, you may or may not see a message on the screen.
VENTILATOR SERVICE REQUIRED
This alarm occurs when the system self-test indicates an error that does not affect the ability of the
ventilator to meet essential performance criteria.
Contact Philips Respironics Customer Service
performance
during active
The device continues to operate (possibly in a reduced capacity mode). If the problem is not corrected,
the device will generate a reminder message until the issue is corrected. Additionally, if therapy is
stopped, a reminder message will immediately appear when therapy is turned on again.
OBSTRUCTION
The ventilator detects an obstruction in the patient’s inhalation path, exhalation path, or external flow
sensor. The ventilator detects that the exhalation port is missing.
Check the circuit for kinked or pinched tubing.
• Check the bacteria filter or HME for a blockage or occlusion.
• Ensure that the exhalation port is not occluded or missing.
• Check the external flow sensor for a blockage or occlusion.
Passive, dual limb, active PAP, or active flow circuit
performance
during active
The device will automatically open the active exhalation valve and continue to operate.
summary
Any circuit, inhalation limb: an obstruction is detected when either of the following conditions exist:
• The flow exiting the machine is less than 0.5 lpm for 5 seconds continuously
• The flow exiting the machine during inspiration is less than 1 lpm for 5 seconds, 2 breaths, or 65
seconds for very low breath rates.
Any circuit: The expiratory port is missing and causes an obstruction alarm if the average flow is less
than 1 lpm for 2 breaths or 60 seconds.
Active flow or dual limb circuit: an obstruction is detected when the external flow sensor measures
less than 0.5 lpm for 65 continuous seconds.
HIGH EXPIRATORY PRESSURE
During the expiratory phase, the delivered pressure exceeds the target patient pressure by 5 cmH
2
• Check the circuit for kinked or pinched tubing.
• Ensure that the leak device is not blocked or occluded.
Note: This alarm condition may be due to pinched tubing or the patient having a fast breath rate.