Instructions for Use | Alarms and System Messages
Alarms and System Messages 46
performance
during active
The alarm is automatically resolved when the delivered pressure comes within 5 cmH
2
patient pressure during the expiratory phase.
Device will continue to operate.
HIGH INSPIRATORY PRESSURE
During the inspiratory phase, the delivered pressure exceeds the target patient pressure by 5 cmH
2
Check the patient for:
o Coughing or excessive secretions
o Bronchospasms
o Tracheotomy tube stability
• Check the ventilator for:
o Kinked, pinched, or blocked tubing
o Blocked leak device or blocked exhalation device
Therapy mode must be SIMV-VC or A/C-VC
performance
during active
The alarm is automatically resolved when the delivered pressure falls within 5 cmH
2
patient pressure during the inspiratory phase.
The device will automatically cycle to the expiratory phase and continue to operate.
EXTERNAL FLOW SENSOR FAILED
A flow sensor offset or sensor failure is detected when the device is in standby or delivering therapy.
Ensure the sensor and cable are connected properly and are not damaged. Clean if needed. Replace if
necessary.
Active flow or dual limb circuit
performance
during active
The alarm remains active until the flow sensor is replaced or reconnected. If the alarm occurs when
delivering therapy, the device will continue to operate.
If the alarm occurs when the device is in standby, you cannot start therapy.
EXTERNAL FLOW SENSOR CABLE DISCONNECTED
The external flow sensor cable becomes disconnected from the ventilator during active therapy.
Active flow or dual limb circuit is selected and an external flow sensor cable is not connected.
Ensure the sensor and cable are connected properly and are not damaged. Replace if necessary.
Active flow or dual limb circuit
performance
during active
Prohibits starting therapy if the external flow sensor cable is not connected to the ventilator when the
active flow or dual limb circuit is selected
EXTERNAL FLOW SENSOR NOT CONNECTED
The external flow sensor becomes disconnected from the external flow sensor cable during active
therapy.
Ensure the sensor and cable are connected properly and are not damaged. Replace if necessary.
Active flow or dual limb circuit
Device continues to provide therapy at the set breath rate.
Device will not trigger on a patient initiated breath.