Instructions for Use | Alarms and System Messages
Alarms and System Messages 51
LOSS OF CO
2
SIGNAL
2
sensor is reporting invalid data, the sensor signal is lost, or no breaths are detected for
more than 10 seconds while the device is delivering therapy.
Ensure that the EtCO
2
sensor is properly attached to the ventilator and the patient circuit.
2
2
alarm is enabled and the sensor was previously reporting valid
data for 3 continuous seconds..
performance
during active
The alarm is automatically resolved when the EtCO
2
sensor is properly attached to the ventilator and
the patient and the sensor reports data.
Device will continue to operate.
LOSS OF SPO
2
SIGNAL
The oximeter is reporting invalid data or the oximeter is disconnected for more than 10 seconds while
the device is delivering therapy or is in standby.
2
probe is properly attached to the ventilator and the patient.
•
Reposition the probe on the patient, if necessary.
2
2
, low pulse rate, or high pulse rate alarm is enabled and the oximeter was
previously reporting valid data for 3 continuous seconds.
performance
during active
The alarm is automatically resolved when the SpO
2
probe is properly attached to the ventilator and to
the patient and the oximeter reports data for more than 10 seconds.
Device will continue to operate.
CO
2
SENSOR ADAPTER ZERO REQUIRED
The CO
2
sensor requests a zero when the device is delivering therapy.
Reestablish the baseline CO
2
2
Sensor Adapter Zero” procedure in the “Device
Options” chapter.
performance
during active
Device will continue to operate.
CO
2
monitored parameters and alarms will not function
CHECK/CHANGE CO
2
AIRWAY ADAPTER
The CO
2
sensor reports that a check is required when the device is delivering therapy.
performance
during active
Device will continue to operate.
CO
2
monitored parameters and alarms will not function
CO
2
SENSOR FAILURE
The CO
2
sensor reports a fault while the device is delivering therapy.
Disconnect and then reconnect the sensor.
Device will continue to operate.
CO
2
monitored parameters and alarms will not function