Page 7.8
SECTION 7 - MAINTENANCE INFORMATION
PXI/PXIe LVDT/RVDT/Resolver Simulator Module 41/43-670
pickering
Table 7.7 - Linearity Verication Test Results For Resolver Simulator -xxxx01
Serial Number:
Channel
Number:
Date of Test:
Operator’s
Name:
Test Equip-
ment Used:
Position
Measured
Sin
Minimum Maximum
Measured
Cos
Minimum Maximum
Result (delete
as appropriate)
-180.00 0.000 0.009 1.782 1.818
Pass / Fail
-170.00 0.302 0.323 1.755 1.791
Pass / Fail
-160.00 0.604 0.628 1.674 1.709
Pass / Fail
-150.00 0.887 0.914 1.542 1.576
Pass / Fail
-140.00 1.142 1.172 1.363 1.395
Pass / Fail
-130.00 1.363 1.395 1.142 1.172
Pass / Fail
-120.00 1.542 1.576 0.887 0.914
Pass / Fail
-110.00 1.674 1.709 0.604 0.628
Pass / Fail
-100.00 1.755 1.791 0.302 0.323
Pass / Fail
-90.00 1.782 1.818 0.000 0.009
Pass / Fail
-80.00 1.755 1.791 0.302 0.323
Pass / Fail
-70.00 1.674 1.709 0.604 0.628
Pass / Fail
-60.00 1.542 1.576 0.887 0.914
Pass / Fail
-50.00 1.363 1.395 1.142 1.172
Pass / Fail
-40.00 1.142 1.172 1.363 1.395
Pass / Fail
-30.00 0.887 0.914 1.542 1.576
Pass / Fail
-20.00 0.604 0.628 1.674 1.709
Pass / Fail
-10.00 0.302 0.323 1.755 1.791
Pass / Fail
0.00 0.000 0.009 1.782 1.818
Pass / Fail
10.00 0.302 0.323 1.755 1.791
Pass / Fail
20.00 0.604 0.628 1.674 1.709
Pass / Fail
30.00 0.887 0.914 1.542 1.576
Pass / Fail
40.00 1.142 1.172 1.363 1.395
Pass / Fail
50.00 1.363 1.395 1.142 1.172
Pass / Fail
60.00 1.542 1.576 0.887 0.914
Pass / Fail
70.00 1.674 1.709 0.604 0.628
Pass / Fail
80.00 1.755 1.791 0.302 0.323
Pass / Fail
90.00 1.782 1.818 0.000 0.009
Pass / Fail
100.00 1.755 1.791 0.302 0.323
Pass / Fail
110.00 1.674 1.709 0.604 0.628
Pass / Fail
120.00 1.542 1.576 0.887 0.914
Pass / Fail
130.00 1.363 1.395 1.142 1.172
Pass / Fail
140.00 1.142 1.172 1.363 1.395
Pass / Fail
150.00 0.887 0.914 1.542 1.576
Pass / Fail
160.00 0.604 0.628 1.674 1.709
Pass / Fail
170.00 0.302 0.323 1.755 1.791
Pass / Fail
180.00 0.000 0.009 1.782 1.818
Pass / Fail
Note: The maximum and minimum values listed do not take into account test equipment uncertainty, values are provided for
convenience only. Only a high performance measuring device should be used to minimize the additional uncertainty.