Why are the measurements incorrect on my Pilot Communications Measuring Instruments?
- YyloweAug 12, 2025
If the measurements on your Pilot Communications Measuring Instruments are incorrect, several factors could be at play: * **Incorrect Voltage Measurement:** Ensure the neutral connection is reliable, the measured voltage matches the equipment's rated parameters, and the PT variable ratio parameter is set correctly. * **Incorrect Current Measurement:** Verify the measured current aligns with the equipment's rated parameters and that the CT variable ratio parameter setting is correct. * **Incorrect Power Measurement:** Confirm the measurement mode settings are correct, the voltage and current phase sequence is correct, and the current namesake terminal is not incorrectly connected.