FR 46 52 0A nn d0..d255 Read 256 byte flash page. nn = page
address in decimal. data returned as
256 binary bytes, with no terminator.
Note : if this command is used while
logging is enabled, spurious results
may be returned if the flash is busy
with a write. Logging should be
stopped (e.g. using the LS
command) before attempting to
download log data.
LS 4C 53 0A nnn <LF> Returns the number of log sessions
in memory. Also stops logging if
currently active
LS i.e. 1st
session:
4C 53 31 0A
nn aaa <LF> Returns flash page address of the
start of session nn. (first session is
1).
Setting n=0 returns the first unused
flash page
Note : this command may take up to
1 sec to complee due to time taken to
search through flash, depending on
flash size and content.
Thermometer operation will be
suspended during this time.
i.e. “3”
OS 4F 53 2D 31
2E 30 30 30
0A
+/-n.nnn OK<LF> Offset Setting in °C
a positive, no sign
Range -3.000 to 3.000
Negative: minus (-) sign
Three digits
i.e. 0.001 will add 0.001 degree
i.e. -0.201 will subtract 0.201 degree
OV 4F 56 0A The offset value will be displayed °C
a positive value, no sign
Range -3.000 to 3.000
Negative value: minus (-) sign
Three digits
i.e. 0.001 will add 0.001 degree
i.e. -0.201 will subtract 0.201 degree