9 RS232 commands
Cmd
Hex Parameters Reply
ID &h 49 44 0A nnnnn <LF> Returns type
i.e. “RH02
TV &h 54 56 0A [0] nnn.nnn[C|F] <LF> Read Temperature.
O : if present, omits C/F suffix
i.e. “20.500C
&h 32 30 2E 30 30 30 43 0A
TV0 &h 54 56 30
0A
i.e. “23.943071363
&h 32 33 2E 39 34 33 30 37
31 33 36 33 0A
SC &h 53 43 0A week;year;batch Serial Code
week : kalender week
Year : a=0, b=1, c=2, d=3 etc
index : 1..99
i.e. week 18, 2026, #01
26CG01
&h 32 36 43 39 30 38
ER aa dd <LF> read eeprom byte. aa = address
(ascii decimal), dd = result (dec)
EW aa,dd OK <LF> write eeprom byte. aa,dd =
address,data (ascii decimal)
RA nnnn <LF> Read raw ADC value. Max =
8388607
i.e. “3855634
FV 46 56 0A version text <LF> Read firmware version
i.e. “Tamtherm 4.05
RT 52 54 0A dd:mm:yy,hh:mm:ss
To write time to the TT3
i.e. “29/03/2025,00:43:11
is written as
29 03 26 00 43 11
response is "OK"
&h 52 54 32 39 30 33 32 36
30 30 34 33 31 31 0A
Values are not range
-
checked
RT 52 54 0A dd:mm:yy,hh:mm:ss To read time from the TT3
i.e. time is:
29/03/2025,00:43:11
TT3 output is
29/03/23,0:52:20
or
&h32 39 2f 33 2f 32 36 2C
39 3A 35 32 3A 32 30 0D
Leading zero
s are skipped
in output from TT3