10.3 Change the interval
When logging with an one second interval this is
60 points per minutes
3.600 per hour
32.000/3.600 = 9 hours
The table shows the total period available for logging using
other sampling intervals
Sample time
Seconds
hours of registration
1 9
5 45
10 91
30 273
60 546
Clear memory with the menu option “Clear Logs”
Connect the thermometer to the mains socket (=battery
charger). When logging for multiple hours or longer