7. CARE AND MAINTENANCE
CARE AND MAINTENANCE
Clean the geneO
+
U at least once a week. Turn the system OFF and wipe down all surfaces with a soft, damp, non-abrasive cloth. Be
careful not to spill any liquids on the geneO
+
U.
Ensure that water never enters the inside of the system. Make sure the system is completely dry before next use.
After every treatment using the Massager Head, clean and disinfect the Massager Head according to the following procedure:
1. Disconnect the Massager Head silicone top cover from its plastic base.
2. Clean with water and detergent solution until it is visually clean.
3. Use a soft, dry cloth to dry the Head.
4. Use a soft cloth, moistened with 70% alcohol available in local pharmacies, to disinfect the Head for at least 1 minute.
5. Use a soft, dry cloth to dry the Massager Head.
The Capsugen is for single use. Care and Maintenance are not required.