(,QWHUIDFH2SHUDWLRQ56DQ G,(((*3,%
( 0HDVXUHPHQW5DQJH6HWWLQJDQG4XHU\
All transmissions from and to the controller are terminated with the ASCII
character LF (Line Feed) i.e. decimal 10, hexadecimal 0x0a. In the examples
given here, this character is described with the escape sequence "\n" accord-
ing to its presentation in the programming language C.
Invalid settings are ignored.
9HORFL W\'HFRGHU
Examples:
Command Description Answer
Answer with
ECHOON
VELO? Queries the measurement range
set
Range number
(ASCII character)
1...9
(refer to table below)
VELO1...
VELO9
VELO1...
VELO9
Sets the corresponding measure-
ment range (refer to table below)
- VELO1...
VELO9
VELO Loads the initialization setting for
the measurement range
(refer to command RENDCL)
- VELO
OVR Queries whether the full scale
range has been exceeded
(Overrange)
0 = no overrange
1 = overrange
OVR0
OVR1
Measurement range Range number
OVD-01 OVD-02
15-
516
25 2 7
125 3 8
1,000 4 9
9(/2? Q
If an OVD-01 is installed, the measurement range is set to
25 .
9(/2? Q
If an OVD-02 is installed, this velocity decoder is selected and
the measurement range is set to 25 .
9(/2"? Q
As an answer, a string with the number of the velocity measure-
ment range set is returned.
If the measurement range 25 of the OVD-02 is selected,
the controller returns the string "7\n".
mm
s
----------
V
mm
s
----------
V
mm
s
----------
V
mm
s
----------
V