Scroll Medical & Laboratory Compressed Air Systems
Please read and save these instrucons. Read carefully before aempng to assemble, install, operate or maintain the product described.
Protect yourself and others by observing all safety informaon. Failure to comply with instrucons could result in personal injury and/or
property damage! Retain instrucons for future reference.
Descripon
Powerex 150 Producon Drive Harrison, OH 45030 USA
P 1.888.769.7979 F 513.367.3125 www.powerexinc.com
IN907800AV 08/2018
Page 1 of 40
The Powerex Scroll Medical Compressed Air Source Systems are
intended to provide Medical Air as where risk meets the Category
3 requirements in NFPA99. Medical Air is intended for use in human
respiraon. Standard Laboratory Compressed Air Source systems are
intended for use in facilies where the risk is less than Category 3,
or applicaons outside of the scope for NFPA99, and not intended
for human respiraon. The systems include intake ltraon, electric
motor driven compressor pumps, aercoolers, condensate traps,
air storage receiver with automac condensate drain, dryer systems
with ltraon and bypass valves, dew point monitor, CO monitor if
applicable, and controls to operate the system.
Safety Guidelines
A SEPARATE SAFETY BOOKLET IS PROVIDED ALONG WITH THIS
MANUAL. READ AND UNDERSTAND THE SAFETY BOOKLET. This
manual contains informaon that is very important to know and
understand. This informaon is provided for SAFETY and to PREVENT
EQUIPMENT PROBLEMS. To help recognize this informaon, observe
the following symbols. MAKE SURE EVERYONE OPERATING OR
SERVICING THE COMPRESSOR READS AND UNDERSTANDS ALL THE
INFORMATION PROVIDED.
Danger indicates an imminently hazardous
situaon which, if not avoided, WILL result in death
or serious injury.
Warning indicates a potenally hazardous
situaon which, if not avoided, COULD result in
death or serious injury.
Cauon indicates a potenally hazardous situaon
which, if not avoided, MAY result in minor or
moderate injury.
Noce indicates important informaon, that if not
followed, may cause damage to equipment.
NOTE: Note indicates informaon that requires special aenon.
Unpacking
Immediately upon receipt of the vacuum system, inspect for any
damage which may have occurred during shipment. Repair or replace
damaged items before use. The name-plate should be checked to
verify the correct model and voltage.
Do not operate unit if damaged during shipping,
handling or use. Damage may result in unsafe
condions and cause injury or property damage.
Compressed Air Type Oilless air meeng requirements of Type 0
Powerex Oilless Scroll compressor pumps
Medical Systems:
- Final ltraon to 0.5 micron or beer, regulated
for 50-55 psig
- Dew point alarm: High dew point at 35°F
- CO Alarm: High CO at 10 ppm
- Dryer purge saving control on at -15°F; o at
-10°F
Laboratory Systems:
- Final ltraon to 0.5 micron or beer, regulated
for 90-95 psig
- Dew point alarm: High dew point at -20°F
(Standard)
- CO alarm (oponal at 10 ppm)
- Dryer purge saving control on at -45°F; o at
–40°F
Lubricaon: Grease lled bearings, PTFE compound p seals
Maximum System
Pressure
Medical Systems: 116 psig
Laboratory Systems: 145 psig
Motor Protecon
System
Motor protector circuit breakers
Air Receiver ASME Code, rated 165 psig or higher
Control Panel UL508 Compliant
Misc. Cal. OSHA 462 (L) (2) Compliant
Cal. OSHPD OSP – 0380-10
Specicaons

Questions and Answers: