EasyManua.ls Logo

Pressure Systems 9022 - Page 119

Pressure Systems 9022
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
www.PressureSystems.com
4 - 9
Following is a simple, step-by-step procedure for executing a specified-value span calibration of
a 9016, 9021, or 9022 Intelligent Pressure Scanner. For Models 9021/9022, skip the steps that
manipulate valves as denoted by the double asterisk (**). For the purposes of this example, it will
be assumed that an upscale pressure of 14.9800 psi is available from a dead weight tester for the
calibration of 15 psi internal transducers. All sixteen channels are 15 psi full scale.
Description 9016 Command 9021/9022 Command
Ensure that valves are in RUN/CAL mode
(default)
[w1200] **
… normal data acquisition
Perform Re-zero calibration first
See Section 4.2.2
Place the module calibration manifolds into
the CAL position if this is the desired
pressure application input. The RUN position
may be a better choice for modules with
transducers having different ranges.
[w0C01] for
CAL
[w0C00] for
RUN
**
With a deadweight tester, apply 14.9800psi
to the appropriate module CAL and CAL
REF inputs [or optionally to RUN inputs]
Delay for settling of pneumatic inputs
Verify measured data reads near expected full
scale
[rFFFF0] [r0FFF0]
Instruct module to calculate new gain
coefficients for all 16 channels
ZFFFF 14.98 Z0FFF 14.98
Place calibration manifold back into the RUN
position
w0C00 **
Store new gain coefficients to transducer non-
volatile memory
w09
… continue normal data acquisition

Table of Contents

Related product manuals