EasyManua.ls Logo

Pressure Systems 9022 - Page 53

Pressure Systems 9022
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
www.PressureSystems.com
3 - 23
Command ‘C’— Sub-command Index 03: Abort Multi-Point Calibration
This sub-command has no additional parameters.
Command C 03
‘C’ is the command letter.
‘ 03’ is the sub-command index (ii) for Abort.
NOTE: all parameters are separated by a space.
Response A
‘A’ is the acknowledge letter
Description: Aborts the Multi-Point Calibration function, if it is currently in progress. This
sub-command also restores the original “A/D samples for averaging” count to the
module that was in use before the calibration function was started.
It should be noted that executing the Configure & Start (‘00’) sub-command again,
after the calibration function has started collecting data (per Collect Data (‘01’) sub-
commands), but before the final data are calculated (per Calculate & Apply (‘02’)
sub-command), will have the same affect as this Abort function.
Example:
! Abort the Multi-Point Calibration function previously started
C 03
Read response:
“A”

Table of Contents

Related product manuals