Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
www.PressureSystems.com
3 - 54
Example:
! Send TCP/IP command to 9016 module (via its connected socket) that returns
decimal voltage data (of the raw temperature signal) for channels 1, 5, 9, and 13:
n11110
Response contains data for channels 13, 9, 5, and 1 (left to right):
0.53013 0.541698 0.503633 0.000000
In this example channels 13, 9, and 5 return normal temperature voltage signals in
the range of 0.5 to 0.6 volts. Note that channel 1 returns a value of 0.0 volts,
indicating a possible error in its temperature signal.
A Model 9021/9022 example would be similar, but without specifying the non-
existent channel 13. (“n01110”)