Regression Statistics: The display will present the linear regression statistics of correlation coefficient
CC and standard error SE. Move down.
Moisture= 12.89%
CC:0.987 SE: 0.04
Calculated Calibration Settings: Using the regression statistics, the software will calculate the correct
calibration settings for the data entered. Move down from the zero screen to view the Span value.
Moisture= 12.89%
New Zero: -23.56
Moisture: 5.1% 01
New Span: 15.00
Moisture= 12.89%
Save? :Prod A (01)
Move Down to Save the NEW
values.
Move the cursor under the digit, then
UP/DOWN to set the desired value.
Press Enter to Save the settings.
Press the LEFT arrow once to return
to the CAL Routine selection, twice
to go Home.