Analog Outputs
The MCT sensor provides a 4-20mA output signal for the temperature reading. Even though the temperature sensor
is set up to measure over 20-120C (70-250F), the 4-20mA signal may be scaled for any desired range within these
limits.
Using the Operator Interface, select the TEMPERATURE menu.
Select CAL PARAMETERS
Select ANALOG HI (2): 100.00
Set the number to the temperature value at which 20mA is desired.
Press ENTER to save the value
Select ANALOG LO (2): 0.0
Set the temperature value at which 4mA is desired.
Press ENTER to save the value.
Return to the HOME screen
Component Parts
P/N Description
240-22 Interface PCB
240-030 IRt/c Temperature Sensor