3302F Series Mainframe Operation Manual 55
4.6.3. STATE functions can be used to see the actual operating status of the electronic
Load at that time.
[STATe
] LOAD {SP}{ONOFF}
Command Syntax:
[STATe] LOAD{SP}{ONOFF}{NL}
[STATe] LOAD {NL}
Purpose: Read LOAD ON/OFF status
Description: This command is used to see if the Load is ON or OFF
0 = Load OFF
1 = Load ON
[STATe
] MODE {SP}{CCCRCVCPLED}
Command Syntax:
[STATe] MODE {SP}{CCCRCVCPLED}{NL}
[STATe] MODE {NL}
Purpose: Set and read the operating mode of LOAD
Description: The return value is 01234 which corresponds to the
Operating mode that the load is in. i.e. CCCRCVCP LED
CC CR CV CP LED
Mode
(value)
(0) (1) (2) (3) (4)
331XF V V V V
333XF V V V V
334XF/G V V V V V
33401F/G V V V V
TABLE 4-9 MODULES FOR EACH SERIES
[STATe
] SHORt {SP}{ONOFF}
Command Syntax:
[STATe] SHORt {SP}{ONOFF}{NL}
[STATe] SHORt {NL}
Purpose: Reads back whether the short circuit test is active or not
Description: 0 = short circuit test active
1 = short circuit test inactive