OPA-102/101
OPERATION MANUAL
- 23 -
SECTION 9. MAINTENANCE and MANAGEMENT
9-1. Lens cleaning
For the proper maintenance and management of the equipment, it is necessary to regularly
clean the lens by cleansing polluted area of the lens.
9-2. Error codes
With the end of probe placed at unclean area, adjusting the span, or some problem with
the equipment, adjusting error is displayed.
(Zero Error) (Span Error)
After eradicating the cause of the error, pressing [SET] key, it will start re-adjusting.
To continuously perform disregarding the error, just press [ESC] key.
(This time, measured value is not precise because of not adjusted)
(CAL error)
It is displayed when CAL and measurement are performed at the same time.
To continuously perform disregarding the error, just press [ESC] key.
(rpm error)
It is displayed when the terminal of rpm at side of the equipment is not connected. Make
sure the terminal of rpm is connected.