Chapter 3 Calibration
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC 3-71
Table 3-13. AEC Linearity Data
SAMPLE 1 mA
(From Table 3-9)
= _______
SAMPLE 2 mA
(From Table 3-9)
= _______
EXPOSURE mR mAs EXPOSURE mR mAs
11
22
33
44
55
66
77
88
99
10 10
SUM SUM
MEAN VALUES
= SUM ÷ 10
MEAN VALUES
= SUM ÷ 10
X
1
= mR ÷ mAs
_______ ÷ _______
X
2
= mR ÷ mAs
_______ ÷ _______
X
1
= _______ X
2
= _______
X
1
- X
2
= _______ X
1
+ X
2
= _______
LINEARITY =
= ________________ must be <
0.1
Page 3 of 4