Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging 2-39
Interfacing to Pausch Tomography System (When Applicable)
To interface the HF Series generator to Pausch tomo system, two lengths of
10-conductor shielded cable is required (supplied with the HF Series genera-
tor). No wire terminations are required for interconnecting the cables. Refer
to Figure 2-22 for Pausch tomo/generator interface wiring. After interfacing,
secure the tomo interface cable securely using cable ties as required.
Check that jumper JP1 is not installed on Universal AEC Board A11
(AY40-027S). If installed, remove this jumper.
IMPORTANT! Be sure to re-program the tomo
sweep times on the Pausch tomography system to
correspond with available time settings selectable
at the HF Series generator Operator Control
Panel.