Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC
5-21
Invalid
AEC
Exposure
Value
Improper AEC calibration. Re-calibrate AEC.
Invalid
AEC
License
1. System is not configured with
AEC option.
2. The AEC license is not active
1. Check AEC configuration settings
are correct.
2. Call Technical Service/Support.
Invalid
License -
Invalid
base
license.
System
Key: {text
string}
Invalid Q-Vision license. Note:
{text string} will contain additional
details.
Call Technical Service/Support.
Invalid
License -
File not
found.
System
Key: {text
string}
Q-Vision license file missing.
Note: {text string} will contain addi-
tional details.
Call Technical Service/Support.
Invalid
Boost
Time
Exposure
Value
The database is corrupt. Call Technical Service/Support.
Invalid
Break
Time
Exposure
Value
The database is corrupt. Call Technical Service/Support.
Invalid
Digital
License
The Digital License is not active. Call Technical Service/Support.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action