Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC
5-43
Save
Configura-
tion
Cable connections to Generator
check out, but configuration data
is corrupt.
1. Power Generator OFF then ON.
2. Use Backup Configuration.
Stuck But-
ton
Button down at power-up. 1. Take finger off button
2. Check keyboard buttons for proper
operation.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action