Chapter 6 Diagrams
HF Series X-ray Generator - Service Manual Revision W
Quantum Medical Imaging, LLC 6-15
Figure 6-8. System Cabling Diagram (with
Q-Connect Option)
4
L1
1
2
L1
SW1
1 2
SW2
1 2
L2/N
L1
L3
L2
L1
L1
L2
1
2
2
BLACK - GND
1
RED +24VDC
3
1
RED +24 VDC
2
BLACK - GND
ETHERNET RJ-45
TABLE BUCKY
WALL BUCKY
TB6
TB3
INTERFACE PCB
TB1
TB1
X-RAY GENERATOR
TB2
ROTOR
PCB *
AY40-013T
AY40-023T
SC50-0 T REV.94 A
SINGLE PHASE OPTION
CATHODE(-)
ANODE(+)
HIGH VOLTAGE
TB1
THREE PHASE OPTION
INCOMING AC POWER
L1 L2 L3
GENERATOR GROUND
N
PHASE A
PHASE B
PHASE C
NEUTRAL
AC Power
115/230 VAC
L1
'X-RAY ON' LAMP
TB5
L2
PHASE A
PHASE B
GROUND
GROUND
TOSHIBA
X-RAY TUBE
ANODE(+)
CATHODE(-)
CATHODE(-)
ANODE(+)
AID - HSS
208-260 VAC
380-480 VAC
or
50-60 Hz
STANDARD
*Note: The Standard Rotor PCB
AY40-013T is removed when the
QG-HSS AID High Speed Starter or the
QMI-HSS is present in the system.
QG - HSS
OPTION *
AY51-017
WALL STAND
J2
ION CHAMBER
TB1
TB1
STANDARD
HANDGRIP
TUBESTAND
DOOR INTERLOCKS
WALL
J2
AEC PCB
TABLE
J4
TB1
TABLE
VARIAN
X-RAY TUBE
TB4
J8
LOGIC PCB
AY40-006S/
OCP INTERFACE PCB
AY40-073S
A1A1P5
COLLIMATOR
WALL BUCKY
COLLIMATOR
TUBESTAND
L1
N
NEUTRAL MAY BE OMITTED
IF NOT REQUIRED BY
LOCAL ELECTRICAL CODE.
AY40-031S
TECHVISION
TB1
GREEN-PIN 2
DA10
BLACK-PIN 3
DUNLEE X-RAY TUBES
CENTER PIN
1300/1400
OUTER PIN
WHITE-PIN 1
INNER PIN
115/230VAC 15A
OPTION
<<<--- CONDUIT --->>>
208-240 VAC
EARTH GROUND
GND
200-240VAC
20A
L1
L2
QMI - HSS
OPTION *
AY40-036S
TB2
COMMON
MAIN
THERMAL SWITCH
THERMAL SWITCH
PHASE
TB4
TS-1 PIN
GREEN-PIN 6
ORANGE-PIN 5
TS-2 PIN
AC Power
TABLE BUCKY
ION CHAMBER
QUANTUM
AUXILIARY HANDGRIP
J3
ETHERNET
SWITCH
GENERATOR
CONTROL BOX
Q-CONNECT
OPTION
WORKSTATION
DISPLAY
COMPUTER
A1A1P1
POWER
AC POWER