Chapter 6 Diagrams
HF Series X-ray Generator - Service Manual Revision W
Quantum Medical Imaging, LLC 6-49
Figure 6-42. Logic Board A1 (AY40-006S) Test Point/
Jumper Location Diagram
AY40-006S REV. K
C20
C
2
2
C23
C26C27
C28
C6
C9
R
1
7
R
1
8
R
1
9
R
2
1
R
2
3
R
2
4
R
2
6
R
3
9
R40
R
4
8
R
5
R53
R54
R59
R60
Y2
R
6
3
R
6
4
R
6
5
R
6
6
C
3
2
C
3
3
C
3
4
C
3
5
C
3
6
C
3
7
C
3
9
C
4
3
C
N
T
R
L
L
O
G
I
C
P
C
B
R
E
V
.
E
A
S
S
Y
A
Y
4
0
-
0
0
6
S
R
E
V
.
S
/
N
1
1
1
4
1
2
9
10
1
1
1
1
1
3
1
2
15
16
J1
J7J6
J9
J
3
TEST
POINT 8
NOTE: “ ” DENOTES JUMPER LOCATION
TEST
POINT 2
TEST
POINT 6
TEST
POINT 5
TEST
POINT 1
TEST
POINT 10
TEST
POINT 4
TEST
POINT 3
TEST
POINT 9
TEST
POINT 7
JUMPER
JP6 (HSS)
CONNECTOR
J8
(SEE TABLE
FOR PIN OUT)
JUMPER
JP7
JUMPER
JP1 (RS-485)
J8-1 HSS ON
J8-2 BOOST ON
J8-3 RUN ON
J8-4 ROTATING
J8-5 TEMP FAULT
J8-6 HIGH/LOW SPEED
J8-7 OVER CURRENT
J8-8 HSS DETECT
CONNECTOR J8 PIN OUT
TP1 - +5 VDC
TP2 - HIGH VOLTAGE ON
TP5 - TABLE BUCKY GRID MOVING
TP6 - WALL BUCKY GRID MOVING
TP8 - OCP COMMUNICATION
TP10 - +4.8 VDC REF ON U14 (A-D)
TEST POINT DESCRIPTIONS
JP1 - ALWAYS INSTALLED
JP6 - INSTALLED ONLY ON
SYSTEMS USING AID HIGH
SPEED STARTER, NOT
INSTALLED ON NON-HIGH
SPEED STARTER SYSTEMS
JP7 - INSTALL ON PINS 2-3
JUMPER CONFIGURATION
J8-1 HSS ON
J8-2 BOOST ON
J8-3 RUN ON
J8-4 ROTATING
J8-5 TEMP FAULT
J8-6 HIGH/LOW SPEED
J8-7 OVER CURRENT
J8-8 HSS DETECT
CONNECTOR J8 PIN OUT
TP1 - +5 VDC
TP2 - HIGH VOLTAGE ON
TP5 - TABLE BUCKY GRID MOVING
TP6 - WALL BUCKY GRID MOVING
TP8 - OCP COMMUNICATION
TP10 - +4.8 VDC REF ON U14 (A-D)
TEST POINT DESCRIPTIONS
JP1 - ALWAYS INSTALLED
JP6 - INSTALLED ONLY ON
SYSTEMS USING AID HIGH
SPEED STARTER, NOT
INSTALLED ON NON-HIGH
SPEED STARTER SYSTEMS
JP7 - INSTALL ON PINS 2-3
JUMPER CONFIGURATION

Table of Contents