A-4 Safety Requirements Cardiac Stress Treadmills User Guide
EMC Declaration Tables
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Cardiac Stress Treadmills are intended for use in the electromagnetic environment specified
below. The customer or the user of the Cardiac Stress Treadmills should assure that they are used
in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Cardiac Stress Treadmills use RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The Cardiac Stress Treadmills are suitable for use in
all establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
NOTE Tests verified with shielded input/output cables only.

Other manuals for Quinton TM55

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quinton TM55 and is the answer not in the manual?

Quinton TM55 Specifications

General IconGeneral
BrandQuinton
ModelTM55
CategoryTreadmill
LanguageEnglish

Related product manuals