C-2 Receiving and Installation Cardiac Stress Treadmills User Guide
the treadmill is connected correctly to the appropriate AC
power source.
installation and operation are in accordance with
standards stated in this manual.
Incorrect installation by unauthorized personnel can lead to
equipment damage and may void the warranty.
Warning! The treadmill must be on a dedicated branch circuit.
No other device should be connected to that circuit.
Excessive risk current (leakage) can result if the
equipment is not properly grounded.
Failure to follow these guidelines will produce a
serious or possibly fatal electrical shock hazard.
Consult a qualified electrician as required.
Site Requirements
The treadmill requires a dedicated AC power line (refer to
“Power Requirements” on page B-2). To ensure electrical
safety, the treadmill is equipped with a three-wire power cord
and three-pronged plug. To maintain ground reliability, the
plug must be connected to an equivalent receptacle.
Caution! The treadmill is designed to operate in a typical
clinical environment with adequate heat dissipation
(1850 Watts maximum).
Place the treadmill on a flat surface, free of moisture and
debris. Maintain a minimum clearance of 1.5 feet (0.5 m) on
each side and 6 feet at the rear.
If the preamp for the Q-Stress system is to be attached to the
handrail, maintain a clearance of 1.5 feet (0.5 m) at the front
of the treadmill.