Page #27 background image
21
54
12Ø
15Ø
16
38
12
2 MeV
Build-up
Cap
Center of
Sensitive
Volume
Dimensions in millimeters
10X6-0.6
Exposure Rate 20 μR/s - 133
R/s
200 nGy/s - 1.17
Gy/s
Dose 100 μR - 589
kR
1 μGy - 5 kGy
Auto Dose
Threshold
22 mR/s 189 μGy/s
Calibration 4% @
60
Co
Rate
Dependence
2%, 10 mR/s - 100 R/s
Energy
Dependence
5%, 40 keV - 1.33 Mev
(with buildup cap)
Construction C552 air-equivalent walls &
electrode, 0.6 cm
3
active volume.
12 m cable
Application High-intensity gamma radiation in-
beam measurements. Irradiators.