EasyManua.ls Logo

Radcal Accu Series - 90 M10 Mas Sensor

Radcal Accu Series
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
90M10 mAs Sensor
Battery-powered, isolated, clamp-on non, invasive mA sensor
Aperature 23mm.
Synchronized to Accu-kV measurement
Real-time waveform.
Automatic zero and no range switching.
2000 mA or 9999 mAs full-scale
Measures absolute value of mA.
Resolution greater of 3-4 digits or .015mA
mA accuracy:
±4% of reading (Limited by 0.6mA RMS noise below 15mA).
mAs accuracy (1-s pulse):
±4% of reading (Limited by 0.6mAs RMS noise below 15ma).
77 μs sample rate.
-2.6% at 720Hz, -3 dB at 2.33 kHz.
Digital waveform sent over the USB port to XLPRO
Automatic turn-on only when kV function is active
Battery lifetime: Approximately 35 operating hours.
Caution: In the presence of strong electromagnetic fields, performance may degrade up to 1 Amp.