90M9 mAs Sensor
Battery-powered, isolated, Invasive mA sensor
Synchronized to Accu-kV measurement
Real-time waveform.
Automatic zero and no range switching.
2000 mA or 9999 mAs full-scale
Measures absolute value of mA.
0.2% mA accuracy
5 μAs or 0.2% mAs accuracy (1-s pulse)
Input impedance: Fixed, resistive, 1 ohm plus fuse and wiring.
Withstands1500 vac between input leads and control unit.
77 μs sample rate
-2.6% at 720Hz, -3 dB at 2.33 kHz.
Digital waveform sent over the USB port to XLPRO
Automatic turn-on only when kV function is active
Battery lifetime: Approximately 30 operating hours.
42