• P1.1 INT. CALIB. – Internal adjustment
Start of internal adjustment process. Adjustment is conducted automatically,
and no user assistance is required. If weighing pan of the balance is loaded,
balance will display a command to unload the weighing pan. If weighing pan is
loaded with a relatively small mass (up to 4% of balance max capacity) then
adjustment process will be carried out automatically, but measuring results
may differ by the amount that was present on balance weighing pan.
or
• P1.2 EXT. CALIB. – External adjustment (with an external weight)
Adjustment with external weight, which value is saved in factory menu of
balance, function unavailable in verified balances.
CAUTION
Function unavailable in verified balances.
• P1.3 USER CALIB. – User adjustment (with an external weight)
Adjustment with external weight at any mass in the balance range, however,
not lower than 30% of max range.
CAUTION
Function unavailable in verified balances.