Switch on continuous transmission in basic measuring unit
Format: C1 CR LF
Accessible answers:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
MASS FRAME
- command understood and in progress
-
response is mass value in basic measuring unit
Frame format
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space
stability
marker
space character mass space unit CR LF
Switch off continuous transmission in basic measuring unit
Format: C0 CR LF
Accessible responses:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and carried out
Switch on continuous transmission in current measuring unit
Format: CU1 CR LF
Accessible responses:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
MASS FRAME
- command understood and in progress
-
response is mass value in current measuring unit
Frame format
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I
stability
marker
space character mass space unit CR LF
Switch off continuous transmission in current weighing unit
Format: CU0 CR LF
Accessible responses:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and carried out