- 89 -
An example 2:
Command:
OMI CR LF
return accessible working modes
Response:
OMI CR LF
2 CR
LF
4
CR LF
12 CR LF
accessible working modes are given in
return:
mode number
OK CR LF
command carried out
Set working mode
Command overview: Command sets particular working mode.
Format: OMS_n CR LF
Response options:
OMS_OK CR LF
command carried out
OMS_E CR LF
error in-course of command execution, no parameter or incorrect
format
OMS_I CR LF
– command understood but not accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command
An example:
Command:
OMS_13 CR LF
set Statistics mode
Response:
OMS_OK CR LF
Statistics mode set
Give current working mode
Command overview: Command returns currently set working mode.
Format: OMG CR LF
Response options:
OMG_n_OK CR LF
command carried out, response: current working mode
OMG_I CR LF
command understood but not accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command.
An example:
Command:
OMG CR LF
return current working mode
Response:
OMG_13_OK CR LF
balance operates in Statistics working mode
Give accessible units
Command overview: Command returns units available for a particular device and for a current
working mode.
Format: UI CR LF
Response options:
UI_”x
1
,x
2
, … x
n
”_OK CR LF
command carried out, returns accessible units
UI_I CR LF
command understood but not accessible at this
moment
x unit symbols, separated by means of comas
x → g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, u1, u2
An example:
Command:
UI CR LF
return available units
Response:
UI_”g, mg, ct”_OK CR LF
response: available units
Set current unit
Command overview: Command sets current unit for a particular device.
Format: US_x CR LF
Response options:
US_ x_OK CR LF
command carried out, response: current unit
US_E CR LF
error in-course of command execution, no parameter or incorrect
format
US_I CR LF
command understood but not accessible at this moment
x parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, msg,
u1, u2, next
CAUTION! if x=next, the command swaps to another avaialable unit on the list (it simulates button
pressing).

Table of Contents