Commands Overview
Zero balance
Format: Z CR LF
Response options:
Z_A CR LF
- command understood and in progress
Z_D CR LF
- command carried out
Z_A CR LF
- command understood and in progress
Z_^ CR LF
- command understood but zeroing range is exceeded
Z_A CR LF
- command understood and in progress
Z_E CR LF
- time limit exceeded while waiting for a stable measurement result
Z_I CR LF
- command understood but not accessible at this moment
Tare balance
Format: T CR LF
Response options:
T_A CR LF
- command understood and in progress
T_D CR LF
- command carried out
T_A CR LF
- command understood and in progress
T_v CR LF
- command understood but taring range exceeded
T_A CR LF
- command understood and in progress
T_E CR LF
- time limit exceeded while waiting for stable measurement result
T_I CR LF
- command understood but not accessible at this moment
Give tare value
Format: OT CR LF
Response: OT_TARE CR LF - command carried out
Response format:
1 2 3 4-12 13 14 15 16 17 18 19
O T Space tare space unit space CR LF
Tare
- 9 characters with right justification
Unit
- 3 characters with left justification
CAUTION!
Tare value is always given in calibration unit.
Set tare
Format: UT_TARE CR LF, where TAREtare value
Response options:
UT_OK CR LF
- command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognised (tare format incorrect)
CAUTION!
Use dot in tare format as decimal point.
Send stable measurement result in a basic measuring unit
Format: S CR LF
Response options:
S_A CR LF
- command understood and in progress
S_E CR LF
- time limit exceeded while waiting for stable measurement result
S_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1 2-3 4 5 6 7-15 16 17 18 19 20 21
S space
stability
marker
space character mass space unit CR LF