- 13 -
Level the balance by turning its feet left/right, follow the pictograms, level marker moves
towards the centre of the circle.
When the marker takes central position, its colour turns from red to green the balance is
correctly levelled.
Go to the home screen.
3.4. MAINTENANCE ACTIVITIES
1. Disassembly a weighing pan and other detachable components (the components differ
depending on a balance type see Unpacking and Installation section). Be careful while
detaching the components so as not to cause any damages to the balance mechanism.
2. Using handheld vacuum cleaner remove dust from the weighing chamber.
3. Using a dry flannel cloth clean glass parts (mild cleanser may be applied if it does not contain
any abrasive substances) for draft shield disassembly instruction go to the next section of this
manual.
4. Using a dry flannel cloth clean disassembled components (mild cleanser may be applied if it
does not contain any abrasive substances).
CAUTION!
Cleaning draft shield while still installed may cause damage of the measuring system.
In order to ease cleaning of glass draft shield panes, it is permissible to remove them following the
below instruction.
Cleaning ABS components:
To clean dry surfaces and avoid smutching use clean non-colouring cloths made of cellulose or
cotton. You can use a solution of water and detergent (soap, dishwashing detergent, glass cleaner).
Gently rub the cleaned surface and let it dry. Repeat cleaning process if needed.
In the case when contamination is hard to remove, e.g. adhesive, rubber, resin, polyurethane foam
residues etc., you can use a special cleaning agents based on a mixture of aliphatic hydrocarbons
that do not dissolve plastics. Before using the cleanser for all surfaces we recommend carrying out
tests. Do not use products containing abrasive substances.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG X2 Series and is the answer not in the manual?

RADWAG X2 Series Specifications

General IconGeneral
BrandRADWAG
SeriesX2 Series
CategoryLaboratory Equipment
DisplayLCD
Data ManagementYes
ConnectivityUSB, RS 232, Ethernet, Wi-Fi
CalibrationExternal
Stabilization Time≤3 seconds
Weighing Unitsg, mg, ct, lb, oz
Power Supply100 - 240 V AC 50/60 Hz
Operating Temperature10°C to 40°C
Pan Size90 mm
Operating Humidity40% to 80% (non-condensing)

Related product manuals